Intravascular Robot Teleoperation Success Rate Drops to 78.1% Under Network Degradation
What happened
Xingyu Chen et al. developed a network simulation and benchmarking framework that integrates intravascular robots with data streams to evaluate the impact of network degradation on remote vascular intervention teleoperation. The study involved 8 clinicians and 8 engineering students, who completed 480 navigation attempts under 15 network conditions. Under the most severe combined degradation conditions, navigation success rate dropped from 93.8% in the reference condition to 78.1%, and average attempt time increased from 63.5 seconds to 94.3 seconds; at a 5% configured packet loss rate, 87.5% of attempts succeeded, but all 8 clinicians considered this condition unacceptable. The framework also supported successful remote navigation over a real network on a second robot platform.
AI 根据报道生成 · 10 days ago更新
Reporting timeline
Follow the reporting to explore different perspectives.
- arXiv cs.RONetwork Emulation and Performance Evaluation Framework for Remote Teleoperation of Endovascular Robots
A research team developed a network emulation and benchmarking framework that combines an endovascular robot with data streaming, for systematically assessing the impact of network degradation on remote endovascular teleoperation. The study involved eight clinicians from interventional neuroradiology and general interventional radiology and eight engineering students, who performed 480 navigation attempts across 15 network conditions, evaluating communication performance, video quality, navigation outcomes, confidence, workload and clinician-rated acceptability.