1.47 mm magnetic soft robot catheter completes in vivo pig pancreaticobiliary duct intervention
What happened
The research team proposed a modular magnetic soft robot catheter with a diameter of 1.47 mm, integrating imaging, fiber Bragg grating (FBG) shape sensing, a flexible gripper, a magnetic anchoring balloon, and a micron-scale ultrasound-triggered release shell into the magnet body, surface, or camera footprint. The authors report that, in in vivo pig experiments, the catheter completed operator-initiated autonomous cannulation of the native papilla without sphincterotomy, and under its own imaging guidance was teleoperated 75 mm into the pancreatic duct, and entered the gallbladder via the cystic duct without a guidewire and imaged its lumen. The team also measured the bending-performance cost of each function: imaging and FBG shape sensing accounted for the main portion; the flexible gripper, magnetic anchoring balloon, and micron-scale ultrasound-triggered release shell together reduced the tip angle under an 8 mT magnetic field only from 70° to 64°. The above in vivo results and performance data are all as reported by the authors.
AI 根据报道生成 · 20 days ago更新
Reporting timeline
Follow the reporting to explore different perspectives.
- arXiv cs.ROModular magnetic soft robotic catheter enables visualized intervention in the pancreatobiliary tree
The research team proposed a 1.47-millimeter modular magnetic soft robotic catheter that integrates imaging, fiber Bragg grating (FBG) shape sensing, a compliant gripper, a magnetic anchoring balloon, and a micrometer-thick ultrasound-triggered release shell into the magnetic body, onto its surface, or onto the camera footprint, and measured the bending-performance cost of each function.